8 Data transfer
68
8.2.1 Examples of the remote control of the instrument
SDTttmmjj
hhmmss
Set date and time (German: (Tag, Monat, Jahr,
Stunde, Minute, Sekunde)
SDTmmddyy
hhmmss
Set date and time (Month, Day, Year, Hour,
Minutes, Seconds)
T (ttt)
Tare or set tare to a specific value
ZERO
Set instrument to 0 (if weight is stable and within
the zero setting range)
Rttt
Adjusts the heating to the required temperature
(30°C ... 230°C)
ROFF
Switch off heating
PWT (ttt.t)
Print weight value and temperature value
Print with time basis in s (ttt.t)
(switch off by transmitting PWT)
NOTE
Each remote control command must terminate with «CR» «LF».
The commands are acknowledged if required.
Input
Description of the function activated
D – – – – –
Five dashes are displayed
DTEST123
tESt123 is displayed
D
The display is dark
T10
-10.000 g (Tare set to = 10 g)
T1
-1.000 g (Tare set to = 1g)
T
Instrument is tared
R100
Adjusts the temperature to 100 °C
Command
Function