
5 Operation
■
45
5.5.3 The change key
•
Changing units
– If drying has already been performed or is under way, the unit of
measurement displayed can be changed using the change key
«
F
». Once a drying procedure is ended, the drying result can be
displayed in all the available units until the start of a new drying
sequence using the change key «
F
».
– Release «
F
» once the unit to which you wish to change is
displayed.
5.5.4 The start/stop key
•
Starting drying manually
– Ensure that there is no drying taking place, i.e. that the instrument
is in weighing mode.
– Press «
START/STOP
» briefly.
– The measurement is started.
•
Stopping drying manually
– Drying is under way.
– Press «
START/STOP
» briefly.
– The measurement is stopped.
•
Statistics Info (only XM 60, XM 60-HR)
– Keep «
START/STOP
» pressed until "STATISTICS INFO" is dis-
played.
– Release «
START/STOP
»
Method : TEST Boost
Dryer setup :
Heat mode
: Boost
Temperature
: 140 C
Timer stop
: off
Time
: 10.0 min
Auto stop
: AdaptStop
Free
: 1/20 D/s
Autostart
: off
Standby Temp
: off
Temperature
: 40 C
Unit
: 100-0%
Weight check
: off
Nominal
: 3.500 g
Upper limit
: 4.000 g
Lower limit
: 2.500 g
Settings for the second method
(all the methods saved are printed)
etc....