5 Operation
28
5.3.6 Balance calibration
For details about calibrating the balance, see chapter 3.8 "Weight
calibration" and see chapter 9.3.1 "Calibrating the balance".
5.3.7 Temperature adjustment
In order to calibrate the temperature, see chapter 9.3.2 "Temperature
adjustment"
5.3.8 Stability
5.3.9 Quick-start
"QUICK-START OFF":
The drying sequence starts as soon as a stable weighing is reached.
"QUICK-START ON":
The drying sequence starts as soon as the «
START
» key is pressed, or
the lid is closed. This enables more accurate moisture analysis of
highly volatile samples.
• SET
BALANCE CAL.
MODE OFF
disabled
EXTERNAL
external
EXT.-DEF.
external with a freely defined
weight (DEF. n.nnn g)
DEF.
0.0000
g
Calibration weight for the
EXT. DEF. mode
• TEMP. ADJMT.
TEMP. CAL.
ON/
OFF
Activate the temperature adjustment
• STABILITY
STABILITY
HIGH
MEDIUM
Setting the stability controle
• QUICK-START
QUICK-START
ON/
OFF
Setting the start condition