6 Determining moisture levels
■
53
6.3.2 Temperature
This function key is used to start the drying temperature input.
The drying temperature can be entered in 1°C increments from
30°C ... 230°C.
After 10 minutes temperatures higher than 200°C are automatically
adjusted down to 200°C over the course of the next 20 minutes.
The temperature setting for drying with moisture analysers is lower
than in the case of drying with the drying oven technique.
6.3.3 Timer stop
This function key is used to define the drying time.
If the timer stop is switched on, the measurement procedure is ended
once the time set has elapsed.
The time can be set in 0.1 min increments from 0.1 min ... 240.0 min.
• DRYING TEMPERATURE
TEMPERATURE
105
°C
Temperature input
• DRYING TIME
STOP TIME
10.0
MIN
Only if the timer stop is switched on
TIMER STOP
ON/
OFF