8 Data transfer
66
8.1
Connection scheme
•
Standard duplex connection
•
Standard duplex connection with additional hardware handshake in
the peripheral instrument
B
Blank (space)
S
Prefix (+,–, space)
DP
Decimal point
D0...D7 Digits
U ...
Unit (only if the weight is stable, otherwise no unit is sent)
CR
Carriage return
LF
Line feed
NOTE
Unused positions are filled with spaces.
The decimal point DP can be between D0 and D7.
moisture analyser RJ 45
D25 / D9
Peripheral instrument
RS 232 out
2
3 / 2
RS 232 in
RS 232 in
6
2 / 3
RS 232 out
GND
5
7 / 5
GND
moisture analyser RJ 45
D25 / D9
Peripheral instrument
RS 232 out
2
3 / 2
RS 232 in
RS 232 in
6
2 / 3
RS 232 out
GND
5
7 / 5
GND
CTS
3
20 / 4 DTR
DTR
7
5 / 8
CTS