8 Data transfer
■
67
•
Pin configuration of the RJ45 socket
8.2
Remote control commands
moisture
analyser
RJ 45
Remark
n.c.
1
Not connected
RS 232 out 2
Out (V24)
CTS
3
In (V24)
VDC
4
Out (9 ... 16V)
GND
5
0V
RS 232 in
6
In (V24)
DTR
7
Out (V24)
EXTBUS
8
In (5V, logic)
Command
Function
ACKn
Acknowledge n=0 off; n= 1 on
CAL
Start calibration (only if EXT is selected)
DN
Reset weight display
D..........
Describe weight display (right-aligned)
@N
Reset info display
@.........
Describe info display
N
Reset instrument
OFF
Switch off instrument
ON
Switch on instrument
PCxxxx
Enter anti-theft code
PDT
Print date and time
PRT
Start printing (Press «
» key)
PST
Start print status
Pn (ttt.t)
Set print mode
n=0 Individually print each value (unstable)
n=1 Individually print a stable value (stable)
n=2 Print after change of load
n=3 Print after each integration period
n=4 Print with time basis in s (ttt.t)