37
status-
bit
Name:
Description
01
Valid actuator attached:
A valid actuator has been recognized. A valid actuator is any digital actuator except nanoX and if the
actuator has a measurement system then it has to be a strain gauge sensor.
02
Drift compensation active:
The drift compensation controller is active.
03
Actuator moving:
The actuator is moving. The first possible cause is the fact that the desired voltage cannot yet be
output because the slew rate limits the alteration speed. The second possible cause is a running
generator
.
04
Function generator running:
A function generator is running.
05
Table-driven generator running:
The table-driven generator is running.
06
High voltage on:
The high voltage output is switched on.
07
reserved
08
reserved
09
reserved
10
reserved
11
reserved
12
reserved
13
reserved
14
reserved
15
reserved
16
reserved
17
reserved
18
reserved
19
reserved
20
reserved
21
reserved
22
reserved
23
reserved
24
reserved
25
reserved
26
reserved
27
reserved
28
Started by power-on:
A software reset has not been performed since the device has been powered on by plugging in the
USB cable.
29
Started by software reset:
A software reset has been performed.
30
High voltage in range:
The high voltage is greater than 135V.
31
Supply voltage in range:
The supply voltage is greater than 20V.
9.3 Error Double Word