phone: +49 (3641) 66880 • fax: +49 (3641) 668866 • [email protected] • http://www.piezojena.com
Appendix
26
Description
Get a
default value
by retrieving it from the device's EEPROM. If the id-parameter is the
identifier of a command then the query fails.
Responses
defp,id,state
on success
nok
on failure
Parameters
Description
Type: Range
id
Identifier of the command or value
W: 0 ... 31
state
Value depending on id-parameter
depending on id-parameter
Errors
error-bit 27, error-bit 28, error-bit 29
Examples
request: "defp,22,1.5"
response: "ok"
Set the response interval of the automatic measurement report to 1.5 seconds.
back to command list
9.1.1.10 Command: hvon
Command
hvon,state
Description
Set the state of the high voltage switch. If the device is not ready (see
status-bit 00
) then the
high voltage cannot be switched on.
Responses
ok
on success
nok
on failure
Query
hvon
Description
Get the state of the high voltage switch.
Responses
hvon,state
on success
nok
on failure
Parameters
Description
Type
Value description
state
Switch state
BOOL
false - off.
true - on.
Errors
error-bit 28, error-bit 29, error-bit 31
Examples
request: "hvon,1"
response: "ok"
back to command list
9.1.1.11 Command: volt
Command
volt,value
Description
Set the desired voltage that should drive the actuator while the drift compensation is inactive.
If the drift compensation is active then the value is stored to not take effect until the drift
compensation is deactivated.
Responses
ok
on success
nok
on failure