NOTICE
Damage due to collisions!
Collisions can damage the hexapod, the load to be moved, and the surroundings.
Make sure that no collisions are possible between the hexapod, the load to be
moved, and the surroundings in the workspace of the hexapod.
Do not place any objects in areas where they can get caught by moving parts.
Stop the motion immediately if a controller malfunction occurs.
INFORMATION
The output voltage of the M850B0448 power supply is decreased when the internal
temperature sensor detects a temperature above 50 °C. As a result the servo mode
of the controller is switched off and the system can no longer be operated in this
state.
1. Ensure that the internal temperature of the power supply does not exceed
50 °C.
−
Place the power supply in a location with adequate ventilation to prevent
internal heat build-up.
2. When the servo mode is switched off automatically due to overheating of the
power supply:
Wait a few minutes to let the power supply cool down.
3. Switch the power supply off and on again.