5 Installation
H-845 Hexapod System
MS222E
Version: 2.1.1
37
NOTICE
Impermissible mechanical load!
An impermissible mechanical load can damage the hexapod.
Only transport the hexapod using a crane. The crane must lift the hexapod by
3 ring bolts and shackles affixed to the motion platform of the hexapod.
To fix the ring bolts and shackles use the three M12 holes in the motion platform
of the hexapod, see Figure 15 on p. 30.
Do
not
hold the hexapod by its struts to position it on a surface.
NOTICE
Warping of the base plate!
Incorrect mounting can warp the base plate. Warping of the base plate reduces the
accuracy.
Mount the hexapod on an even surface. The recommended evenness of the
surface is 100 µm.
Requirements
You have read and understood the General Notes on Installation (p. 33).
If possible: Position the surface where the hexapod is to be mounted
horizontally.
You have connected the hexapod to a crane via 3 ring bolts and shackles as
described in “Unpacking” (p. 27)
Tools and accessories
6 M12x60 screws and Allen wrench 10.0, included in the scope of delivery
(p. 19)
Alternative: 6 M16x60 screws and suitable tool, for rotationally symmetrical
affixing of screws, not included in the scope of delivery