35
SETTING THE SELECTABLE PARAMETERS
Selectable parameters
The setting of some parameters (including the alarm setting) can be changed. Following shows the
selectable parameters. Refer to the following pages for the details.
①
1/10 Temp. Display
The temperature display format on the home screen is selectable; integer display or decimal display.
②
Alarm Setting
The selectable parameters; high temperature alarm and low temperature alarm, delay of door alarm
buzzer, buzzer suspended period, buzzer volume, remote alarm.
③
Interior Light
The setting of interior light; linked with the door operation, always ON or always OFF and setting of
lighting period.
④
Comp. Delay
The delay time of compressor start after recovery form the power failure is selectable.
⑤
DAQ Setting
The selectable parameters; device ID for data acquisition, communication speed of data acquisition,
remote control.
⑥
Date & Time
The selectable parameters; setting of date and time, and format of date/time indication “YY/MM/DD” or
“DD/MM/YY)
①
②
③
④
⑤
⑥
A l a r m s & C o n t r o l s
1/10 Temp. D i s p l a y
A l a r m S e t t i n g
L i g h t S e t t i n g
Comp. Delay
DAQ Setting
Date & Time