45
SETTING THE SELECTABLE PARAMETERS
Setting the compressor delay
The delay time before the compressor starts can be changed to reduce the load on the power line and to
facilitate the start-up (reset) of the refrigerator during recovery after power failure.
<Important>
Change the delay time only when the multiple devices are connected to one power source.
Setting range of delay time: between 5 minutes and 15 minutes
Initial setting (factory setting): 5 minutes
1.
On the home screen, press the menu key (MENU/HOME).
►The Menu screen is displayed.
2.
By pressing the up or down key, move the cursor to the Alarms &
Controls [Fig. 1] and then press the enter key (ENTER).
►The Alarms & Controls screen is displayed.
3.
By pressing the up or down key, move the cursor to the Comp. Delay
[Fig. 2] and then press the enter key (ENTER).
►The Comp. Delay screen is displayed, and the current setting (05) is
displayed [Fig. 3].
4.
By pressing the right or left key, select the digit to be changed and
then change the figure by pressing the up or down key.
5.
Press the enter key (ENTER).
►The setting is memorized and the screen is changed to the Alarms &
Controls screen.
6.
Press the menu key (MENU/HOME) to display the home screen.
The display returns to the home screen automatically when 90
seconds has passed without any key operation.
[Fig. 1]
[Fig. 2]
[Fig. 3]
Comp. Delay
05
min
5min - 15min
Menu
Set Temp.
5
o
C
Keypad Lock
Temp. Max / Min
Data Log
Alarms & Controls
Alarms & Controls
1/10 Temp. Display
Alarm Setting
Light Setting
Comp. Delay
DAQ Setting
Date & Time