REMOTE ALARM TERMINAL
The alarm is relayed to a remote location when a remote alarm device (commercially available) is
connected to the remote alarm terminal. Installation of a remote alarm device is recommended when the
unit is installed in an unattended location so that the operator of the unit is notified of the alarm.
Contact our representative or agent to arrange the installation of a remote alarm device.
Location of remote alarm terminal: left bottom (back side)
Allowable contact capacity: DC 30 V, 2 A
It is recommended to use interface cables with a maximum length of 30 meters.
Contact output
Terminal
At normal
At abnormal
Between COM. and N.O.
open
close
Between COM. and N.C.
close
open
<Important>
The remote alarm terminal will be in alarm mode when the power supply cord is unplugged or the circuit
breaker is turned off because this will be interpreted as a power failure.