DISPLAY AND EXPORT OF DATA LOG
Setting the log interval
The log interval for the chamber temperature (refrigerator and freezer) can be set.
Setting range of log interval: between 1 minute and 15 minutes
Initial setting (factory setting): 1 minute (log for approximately 3 months can be stored)
1.
On the home screen, press the menu key (MENU/HOME).
►The Menu screen is displayed.
2.
By pressing the up or down key, move the cursor to the Data Log [Fig.
1] and then press the enter key (ENTER).
►The Data Log screen is displayed.
3.
By pressing the up or down key, move the cursor to the Data Log
Setting [Fig. 2] and then press the enter key (ENTER).
►The Data Log Setting screen is displayed [Fig. 3].
4.
With the cursor on the Data Log Interval [Fig. 3], press the enter key
(ENTER).
►The Data Log Interval screen is displayed [Fig. 4].
5.
By pressing the right or left key, select the digit to be changed and
then change the figure by pressing the up or down key.
6.
Press the enter key (ENTER).
►The setting is memorized and the screen is changed to the Data Log
Setting screen.
7.
Press the menu key (MENU/HOME) to display the home screen.
The display returns to the home screen automatically when 90
seconds has passed without any key operation.
[Fig. 1]
[Fig. 2]
[Fig. 3]
[Fig. 4]
Menu
Set Temp.
5
o
C
Keypad Lock
Temp. Max / Min
Data Log
Alarms & Controls
Data Log
Data Log Chart
Data Log Export
Data Log Setting
Alarm
Alarm Export
Data Log Setting
Data Log Interval
Unique File ID
Data Log Interval
01
min
1min - 15min