Order
Pa-
ra-
me-
ter
Description
Functions
Min
Max
De-
fault
set-
ting
LEV10
none
Returns the
status of the
operating pa-
rameters de-
fined by SET
Example:
#adr,nnnnn,sssss,00000,0,ccccc,eeeee,00000,0000,0000,j
j,kk,ll,mmm<CR>
nnnnn = nominal speed (min
-1
)
sssss = standby speed (min
-1
)
00000 = not used
0 = not used
ccccc = running time of the pump (hours)
eeeee = running time of the electronic drive
00000 = not used
0000 = not used
jj = ‘At speed’ relay setpoint (3 to 50%)
kk = temperature setpoint (30 to 65°C)
ll = landing bearing warning threshold (0 to 99%)
mmm = landing bearing counter (0 to 100%)
NSP
none
Switches
from Stand-
by speed to
nominal
speed
The speed is set to the nominal speed (default setting)
OPT
XX
Setting of us-
er's options/
orders
Example: #adrOPTXX,n<CR>
● XX = 14 control mode
n = 0 HHR control mode (keyboard)
n = 1 Remote control mode (hardware)*
n = 2 RS-232/RS-485 control mode
n = 5 Profibus fieldbus control mode
n = 8 EtherCAT fieldbus control mode
* not possible with EtherCAT
● XX = 25 Braking control
n = 1 no motor braking
n = 0 motor braking
0
1
0
● XX = 29 Temperature management system ON/OFF
(MT pump version)
n = 0 temperature management disabled
n = 1 temperature management enabled
0
1
0
● XX = 33 contact mode
n = 0 Rotating (contact closed if speed > 120 min
-1
)
n = 1 accelerating (contact closed during acceleration
phase)
0
1
0
RDI
none
Serial num-
ber of the
electronic
drive unit
Serial number, coded with 10 digits and stored in EE-
PROM, of the pump, equipped spindle and control inter-
face serial numbers.
RPM
none
Stand-by
speed setting
Example: #adr,nnnnn<CR>
nnnnn = stand-by speed
15000 50000 15000
SBY
none
Switches
from the se-
lected rota-
tion speed to
the Stand-by
speed
The pump runs at the last Stand-by speed stored value.
This last can be modified with RPM order.
Interfaces for control
53/78