1
2
1 Strap
2
Lifting ring M8
Handling the pump using a lifting device
When handling the pump,
a lifting device appropriate for the weight of the product must be used
.
The weight and center of gravity vary depending on the model (see chapter “Dimensions”). Provide 2
approved M8 lifting rings (supplied by the customer).
1. Screw the lifting rings into 2 of the 3 M8 holes on the inlet housing.
2. Use a multi-strand strap with a minimum length of 230 mm
3. Secure the pump using lifting rings and hoist it.
4.
Always transport
the turbomolecular pump in an upright position.
3.3 Storage
Storing a new pump
► Keep the pump wrapped in its protective envelope.
► Always
leave
the inlet, the exhaust and purge blanking plates in place.
► Always
leave
plastic plugs in place on the electrical connectors.
► Store the pump in line with the permitted storage temperatures (see chapter "Environmental con-
ditions").
► Store the pump in a clean, dry and non-contaminated area,
for a maximum period of 1 year
.
Storage after use
► To store a pump that has been used, see chapter "Shutting down for longer periods".
Transportation and Storage
17/78