Order
Pa-
ra-
me-
ter
Description
Functions
Min
Max
De-
fault
set-
ting
SEL10
none
Status of the
options/
orders set
with OPT or-
der
Example: #adr,0,0,1,0,r<CR>
0 = not used
1 = not used
r = returns the remote con-
trol choice
r = 0 Hand Held Remote
(keyboard)
r = 1 Remote control
r = 2 Serial link
r = 5 Profibus
r = 8 EtherCAT
by default, the value de-
pends on the pump config-
uration
SEL20
none
Status of the
parameters
set with
OPTXX
Returns the status of parameters set with OPTXX: S
25
, S
29
,
S
33
Example : #000,S
25
,S
29
,S
33
<CR>
SEP
none
Separator
character
Valid for parameters returned on DLR, STA and LEV.
ASCII input value of the character
code 044 corresponds with comma “,”
000
255
044
SET
XX
Setting pump
operating pa-
rameters
Example: # adrSETXX,ccccc<CR>
XX = 10 pump running time (in hours)
XX = 30 ‘at speed’ relay setpoint (%)
- 3
- 50
- 3
XX = 31 temperature control (°C) (MT pump version)
30
65
30
XX = 32 landing bearing threshold (%)
0
99
20
ccccc = value
TMP
ON
or
OFF
Start/Stop
the pump
The pump rotation starts with TMPON
The pump stops with TMPOFF
VER
none
Display of
software ver-
sions
#adr,INTERFACE Vx.yy.zz,CARTRIDGE Vx.yy,TYPE
zzzz,FIRMWARE Vx.yy.zz
<CR>
INTERFACE: vx.yy.zz = control interface software version
CARTRIDGE vx.y = variable speed drive software version
Type zzzz = type of connected pump
FIRMWARE Vx.yy.zz = software version
Or-
der
Pa-
ra-
me-
ter
Description
Functions
STA
none
Pump status
Example : #adr,s
1
s
2
s
3
, rrrrr,vvv, www, xxx, yyy, zzz, aa, bbbbb, ccc, ddd,
gggggggggggggggggggggggg<CR>
rrrrr = pump rotation speed (min
-1
)
aa = motor voltage (V)
bbbbb = motor current (mA)
ccc = pump temperature (° C)
ddd = electronic drive unit temperature (°
C)
s
1
s
2
s
3
= requires to convert ASCII to bi-
nary
vvv = Radial Xh
www = Radial Yh
xxx = Radial Xb
yyy = Radial Yb
zzz = Axial Z
g
0
to g
24
= warning and default bytes
Interfaces for control