![PerkinElmer OPERETTA Скачать руководство пользователя страница 59](http://html1.mh-extra.com/html/perkinelmer/operetta/operetta_technical-manual_1530224059.webp)
5.3
Specifications
5.3.1
Objective Lenses
Objective
Numerical
aperture
Working
distance
[mm]
Field of
view [µm
2
]
Depth of
focus
[µm]
Optical
resolution
xy [µm]
2x long WD
0.08
6.2
6754 × 5086
145.5
3.75
10x long WD
0.3
10
1351 × 1017
10.3
1
10x high NA
0.4
3.1
1351 × 1017
5.8
0.75
20x long WD
0.45
6.6-7.8
675 × 509
4.6
0.67
20x high NA
0.75
0.6
675 × 509
1.7
0.4
40x long WD
0.6
2.7-4.0
338 × 254
2.6
0.5
40x high NA
0.95
0.18
338 × 254
1.0
0.32
60x long WD
0.7
1.5-2.2
225 × 170
1.9
0.43
60x high NA
0.9
0.2
225 × 170
1.1
0.33
100x long WD
0.85
1.2-1.9
135 × 102
1.3
0.35
5.3.2
Sample Carrier
Recommended sample carriers for the Operetta:
Plate type
Bottom
material
supports
adherence of
cells
Plate bottom
allows use of
high NA
objectives
Suitable for
the following
objectives
ViewPlate-96 F TC (black, with
lid, TC, sterile, individually
pouched) bottom thickness of
0.8 mm
yes
no
2x long WD
10x long WD
10x high NA*
20x long WD
40x long WD*
ViewPlate-384 F TC (black, with
lid, TC, sterile)
yes
no
2x long WD
10x long WD
10x high NA*
20x long WD
40x long WD*
5 Technical Data
Operetta Technical Manual
59
Содержание OPERETTA
Страница 1: ...OPERETTA Technical Manual...
Страница 5: ...1 System Safety 1 System Safety Operetta Technical Manual 5...
Страница 16: ...1 7 Conformity 1 System Safety 16 Operetta Technical Manual...
Страница 18: ......
Страница 19: ...2 Hardware Description 2 Hardware Description Operetta Technical Manual 19...
Страница 33: ...3 Retooling and Maintenance 3 Retooling and Maintenance Operetta Technical Manual 33...
Страница 44: ...4 IT Policy 4 IT Policy 44 Operetta Technical Manual...
Страница 48: ...4 1 3 3 Router 4 IT Policy 48 Operetta Technical Manual...
Страница 55: ...5 Technical Data 5 Technical Data Operetta Technical Manual 55...
Страница 61: ......