4. Close the sample lid.
5. Confirm the message box and click
OK
.
If the plate type – objective lens combination is not working together, the software will
warn you beforehand.
Objective Lens Correction Collar Possible Values [mm]
2x long WD
No
–
10x long WD
No
–
10x high NA
No
–
20x long WD
Yes
0-2
20x high NA
No
–
40x long WD
Yes
0-2
40x high NA
Yes
0.11-0.23
60x long WD
Yes
0.1-1.3
60x high NA
Yes
0.11-0.23
100x long WD
Yes
0-0.7
3.1.3
Emission Filter
The Operetta is equipped with an eight position emission filter wheel. The four
emission filters 410 - 480 nm, 500 – 550 nm, 560 – 630 nm and 650 – 760 nm are
always included in the instrument. All of the other selections can be chosen as an
option at any time for customer installation. Emission filters are barcoded and can be
exchanged by the user. Unused filters can be stored in the filter hotel behind the door.
Follow the instructions of the
Change Filter
wizard in the Harmony software
(
Settings – Change Filter
).
3 Retooling and Maintenance
38
Operetta Technical Manual
Содержание OPERETTA
Страница 1: ...OPERETTA Technical Manual...
Страница 5: ...1 System Safety 1 System Safety Operetta Technical Manual 5...
Страница 16: ...1 7 Conformity 1 System Safety 16 Operetta Technical Manual...
Страница 18: ......
Страница 19: ...2 Hardware Description 2 Hardware Description Operetta Technical Manual 19...
Страница 33: ...3 Retooling and Maintenance 3 Retooling and Maintenance Operetta Technical Manual 33...
Страница 44: ...4 IT Policy 4 IT Policy 44 Operetta Technical Manual...
Страница 48: ...4 1 3 3 Router 4 IT Policy 48 Operetta Technical Manual...
Страница 55: ...5 Technical Data 5 Technical Data Operetta Technical Manual 55...
Страница 61: ......