The flow of carbon dioxide is only active ...
l
if CO
2
control has been activated in the
TCO Settings
(Harmony software)
and
l
if the Live Cell Chamber is connected so that the sample chamber is sealed
and
l
if the flap for the scanning table is closed (only with automation upgrade).
l
CO
2
tubing:
o
Outer diameter: 6 mm
o
Max. bending radius: 21 mm
o
Shore hardness: D 62 ±3
o
Material: TPE-A (polyamide)
l
Working pressure: 3 bar
l
Required CO
2
flow: > 15 l/h
l
Required CO
2
purity: ≥ 99.998 %, particle-free
Notice
If you connect or disconnect the Operetta from the CO
2
supply, dust
particles may be introduced into the tubing. This can lead to
malfunctions and blockages.
l
Blow through the tubing using CO
2
before reconnecting it to the
Operetta.
2 Hardware Description
30
Operetta Technical Manual