![PerkinElmer OPERETTA Скачать руководство пользователя страница 27](http://html1.mh-extra.com/html/perkinelmer/operetta/operetta_technical-manual_1530224027.webp)
2.3
Temperature and CO
2
Option (TCO)
2.3.1
Overview
If the Operetta is equipped with the TCO upgrade, the climate in the sample chamber
can be controlled via the Harmony software:
l
Temperature:
37 °C or unregulated
l
CO
2
concentration:
3 to 8.5 % (0.5 % steps) or unregulated
Notice
The TCO unit needs approximately one hour to reach and stabilize
the target temperature and 5 min. to stabilize the selected.CO
2
concentration.
Inside the Operetta there is an air-circulating unit which regulates the CO
2
level in the
sample chamber. Temperature is controlled by heating elements in the
Live Cell
Chamber
and inside the sample chamber. The sample chamber is sealed so that
almost no air or CO
2
gas can escape.
1. Live Cell Chamber
2. Electrical connector
3. Temperature sensor
2 Hardware Description
Operetta Technical Manual
27
Содержание OPERETTA
Страница 1: ...OPERETTA Technical Manual...
Страница 5: ...1 System Safety 1 System Safety Operetta Technical Manual 5...
Страница 16: ...1 7 Conformity 1 System Safety 16 Operetta Technical Manual...
Страница 18: ......
Страница 19: ...2 Hardware Description 2 Hardware Description Operetta Technical Manual 19...
Страница 33: ...3 Retooling and Maintenance 3 Retooling and Maintenance Operetta Technical Manual 33...
Страница 44: ...4 IT Policy 4 IT Policy 44 Operetta Technical Manual...
Страница 48: ...4 1 3 3 Router 4 IT Policy 48 Operetta Technical Manual...
Страница 55: ...5 Technical Data 5 Technical Data Operetta Technical Manual 55...
Страница 61: ......