PerkinElmer OPERETTA Скачать руководство пользователя страница 41

3.2.2

Cleaning

Cleaning Schedule

Daily: Clean the housing surfaces (outside and sample room).
Only if required: Clean the microscope objective lens(es) carefully using optical wiper
and 70% ethanol.

Cleaning the Housing Surfaces

Notice

The Operetta is not protected against spraying or even splashing
water! The water could get into the housing and damage optical and
electrical components of the instrument.

l

Do not drop or pour water over the instrument.

l

Switch off the Operetta.

l

Clean the surfaces of the housing and the sample room with a slightly damp
cloth. Soiled areas may also be cleaned using 70% ethanol.

l

The glass panels should be cleaned with a suitable glass cleaning agent.

l

Do not handle with greater amounts of fluids near the device.

Cleaning the Objective Lens

Caution!

Sharp-edged sample carrier holders! Be careful when performing
retooling or maintenance tasks on the scanning table.

l

Do not touch the sample carrier holders.

l

The objective lens(es) should only be cleaned if they are
obviously soiled or dusty.

l

Shut down the Operetta to move the scanning table to the service position.

l

Wet the objective lens with 70% ethanol.

l

Take a dust-free optical tissue, wipe the lens a few times very carefully without
force and let it dry. Do not wipe back and forth but carefully in one direction.

3 Retooling and Maintenance

Operetta Technical Manual

41

Содержание OPERETTA

Страница 1: ...OPERETTA Technical Manual...

Страница 2: ...or translated into any language without the prior written permission of PerkinElmer Inc Copyright 2011 PerkinElmer Inc Printed in the UK Trademarks Registered names trademarks etc used in this documen...

Страница 3: ...nterference 14 1 6 Warranty Information 14 1 7 Conformity 16 1 8 Service Schedule 17 1 9 Setup Requirements 17 2 Hardware Description 19 2 1 Introduction 20 2 1 1 Instrument Features 20 2 2 Operetta D...

Страница 4: ...equirements for Harmony Installation on an Office PC 51 4 4 Computer Systems 52 4 4 1 Network 52 4 4 2 Hardware 52 4 4 3 Operating System Software Software Applications 52 4 4 4 Data 53 4 4 5 Back Up...

Страница 5: ...1 System Safety 1 System Safety Operetta Technical Manual 5...

Страница 6: ...r for usage in biological medical and pre clinical research biosafety levels 1 and 2 see also section 1 3 3 Process Liquids or Substances page 9 Due to its technical design the Operetta is particularl...

Страница 7: ...inor injury Notice Failure to observe may result in invalid measurement results or damage of the instrument Specific symbols are used which show you the type of hazard Biohazard Risk of crushing Failu...

Страница 8: ...ance procedures described in this manual are excepted and may be carried out by the user l The manufacturer can accept no liability for this product if the provisions of this documentation are not com...

Страница 9: ...environment BSL 2 is suitable for work involving agents of moderate potential hazard to personnel and the environment All procedures performed on the Operetta should be done carefully to minimize the...

Страница 10: ...that require access to the instrument while the interlocks are defeated may only be carried out by qualified service personnel Service personnel must observe the instructions of the separate Service M...

Страница 11: ...w into the instrument This corresponds approximately to the amount of carbon dioxide which one exhales during one hour and can be considered as relatively safe However the flow limiter does not protec...

Страница 12: ...the provided power cable l Check the power cable for damages before switching on the Operetta Do not use the instrument with damaged power cable l In case of a damaged power cable contact PerkinElmer...

Страница 13: ...ing the outer surface of the Operetta use a slightly moistened damp cloth For cleaning the workstation inside the operator must specify suitable precautions especially with regard to biological contam...

Страница 14: ...cause harmful interference to radio or television reception which can be determined by turning the equipment off and on the user is encouraged to try to correct the interference by one or more of the...

Страница 15: ...erference with the equipment including but not limited to the removal or exchange of parts and the use of accessories of other origin Any such action or negligence will invalidate any claim under warr...

Страница 16: ...1 7 Conformity 1 System Safety 16 Operetta Technical Manual...

Страница 17: ...vibrations Blurred images can be the consequence l Put the Operetta on a very stable bench or table l Do not put other devices on the same table which cause vibrations e g a centrifuge l Avoid any sho...

Страница 18: ......

Страница 19: ...2 Hardware Description 2 Hardware Description Operetta Technical Manual 19...

Страница 20: ...entials of setting up an HCS assay and how to use the Operetta and image analysis software for the most common applications 2 1 1 Instrument Features l High content imaging reader l Switchable between...

Страница 21: ...ot installed and if you reach into the danger area of table or robot l Do not use an Operetta without protective shields installed if the scanning table flap is unlocked l The table in out functionali...

Страница 22: ...Barcode Reader Port for connecting a PerkinElmer barcode reader Notice Please use a PerkinElmer barcode reader only otherwise a faultless operation cannot be guaranteed 2 2 3 Front Side 1 Filter acce...

Страница 23: ...on prior permission from Harmony o Exchange sample carrier Use the Open Lid button in Harmony For details please see Operetta Operation Manual section 3 15 Open Lid o Exchange or clean objective Use t...

Страница 24: ...nt is in operation measuring The LEDs indicate the progress of the measurement on a percentage basis Blinking Blue Red Waiting for User The instrument is waiting for user interaction This does not ind...

Страница 25: ...pen the sample lid In contrast to the status light this is no status indication see also previous chapter 2 Scanning stage with plate holder o Mechanism to hold or release the sample carrier o Sample...

Страница 26: ...ctives 2x 10x 20x 40x 60x 100x long WD and high NA 4 Transmission light source for brightfield excitation o Detect cells independent of dye label free techniques o 740nm Infrared to minimize cell dama...

Страница 27: ...one hour to reach and stabilize the target temperature and 5 min to stabilize the selected CO2 concentration Inside the Operetta there is an air circulating unit which regulates the CO2 level in the s...

Страница 28: ...in the sample chamber is lost and you have to wait until the target values have been reached and stabilized again Inside the Live Cell Chamber there are heating elements for controlling the temperatur...

Страница 29: ...nger Carbon Dioxide Gas Supply Tubing l Do not expose the tubing to any mechanical or thermal stress l Comply with the maximum bending radius of the tubing see next page l If the tubing has to be laid...

Страница 30: ...ng o Outer diameter 6 mm o Max bending radius 21 mm o Shore hardness D 62 3 o Material TPE A polyamide l Working pressure 3 bar l Required CO2 flow 15 l h l Required CO2 purity 99 998 particle free No...

Страница 31: ...y settings on the Harmony PC Please see chapter 4 IT Policy page 44 2 5 Xenon Light Source The Operetta is equipped with an external Xenon light source Notice This chapter is an overview for the most...

Страница 32: ...t Accepts ACMI Olympus Storz or Wolf fiber optic light cables The names of the ports are clearly visible on the perimeter of the adapter outboard from the port inputs Operetta ACMI Danger Danger of wh...

Страница 33: ...3 Retooling and Maintenance 3 Retooling and Maintenance Operetta Technical Manual 33...

Страница 34: ...closely and observe both the general safety instructions see chapter 1 System Safety page 5 and any specific safety instructions which may be given at the beginning of each section 3 1 Retooling 3 1 1...

Страница 35: ...n the glass plate inside the Operetta for 30 60 minutes Follow the instruction of the Objective Exchange Wizard Harmony Settings Change Objective 1 If all four positions of the turret are occupied Sel...

Страница 36: ...pe will also be stated Also please note that while the message box Please adjust the correction collar is displayed the table is positioned so as to give easy access to the objective The objectives ha...

Страница 37: ...m adjustment is shown How to Adjust the Correction Collar Objective with correction collar 1 If the message Please adjust the correction collar to mm appears open the sample lid 2 Turn the correction...

Страница 38: ...long WD Yes 0 0 7 3 1 3 Emission Filter The Operetta is equipped with an eight position emission filter wheel The four emission filters 410 480 nm 500 550 nm 560 630 nm and 650 760 nm are always incl...

Страница 39: ...can damage the instrument 1 Open the filter access door to reach the filter wheel 2 If necessary remove the old filter 3 Insert the new filter to the intended position 4 Close the door The filter whee...

Страница 40: ...symptoms indicating the need for replacement are l Repeated ignition attempts without lamp going into operation l Low light output evident with dark picture at high light level adjustment The deliver...

Страница 41: ...be cleaned using 70 ethanol l The glass panels should be cleaned with a suitable glass cleaning agent l Do not handle with greater amounts of fluids near the device Cleaning the Objective Lens Caution...

Страница 42: ...fective fuse may only be replaced by exactly the same part from the following manufacturer LITTLEFUSE 218000 Series 5 20 mm Time Lag Fuse Slo Blo 15 A 250 V US 8 A 250 V EU 1 Switch off the Operetta a...

Страница 43: ...rried out as specified to guarantee the correct and reliable operation of the Operetta Warning The jobs listed here may only be performed by the PerkinElmer Service l If you do not have a maintenance...

Страница 44: ...4 IT Policy 4 IT Policy 44 Operetta Technical Manual...

Страница 45: ...ible for problems caused by violating the following policies Any effort required to verify this type of problem is billable at the current service rate and is not covered by guarantee and or service c...

Страница 46: ...ion General Network Requirements l The subnet mask for all network adapters has to be 255 255 255 0 l All network connections must be Gigabit Ethernet connections using CAT 5e cable All devices shown...

Страница 47: ...llowed in the customer s company network use one of the following solutions to isolate the Operetta network from the company network 4 1 3 1 Office PC as interface small solution 4 1 3 2 Office PC as...

Страница 48: ...4 1 3 3 Router 4 IT Policy 48 Operetta Technical Manual...

Страница 49: ...rated into a domain l Anti virus or malware protection software may not be installed on this computer 4 1 4 1 Standard Configuration This scenario is only possible if NIC2 of the Harmony PC and NIC1 o...

Страница 50: ...ervice Pack 3 only for office PC installations This includes all security patches until October 1st 2010 Notice PerkinElmer cannot guarantee that future Windows updates provided by Microsoft will not...

Страница 51: ...current service rate PerkinElmer may not provide maintenance service on the computers supplied by the customer Basic l Harmony Office License incl USB dongle with valid Software Maintenance Agreement...

Страница 52: ...hould be discussed in advance with PerkinElmer service to assess the detailed specification Problems with hardware should also be reported to PerkinElmer service 1 The relocation of the Harmony PC is...

Страница 53: ...ony software is tested with Microsoft Forefront Client Security 1 PerkinElmer is not responsible for the implementation of an effective virus security strategy It is suggested to exclude the following...

Страница 54: ...manually must be accessible for the default user account Operetta read write access including the right to create subfolders 4 Passwords Operetta System users should not change any passwords Any chan...

Страница 55: ...5 Technical Data 5 Technical Data Operetta Technical Manual 55...

Страница 56: ...The illumination intensity in the sample can be tuned in the software using a mechanical attenuator The fluorescence excitation wavelength is selected by choosing an excitation filter on an eight pos...

Страница 57: ...citation filters four dichroics four emission filters 20x objectives camera Xenon lamp laser based auto focus and one set of Harmony Software included in the instrument Please see the following table...

Страница 58: ...axis Plate holder Automation plate handler II optional 5 2 1 Detection Components Detector CCD camera with 1 3 mega pixel Peltier cooled 14 bit resolution Emission band pass changer 5 2 2 Light Source...

Страница 59: ...WD 0 7 1 5 2 2 225 170 1 9 0 43 60x high NA 0 9 0 2 225 170 1 1 0 33 100x long WD 0 85 1 2 1 9 135 102 1 3 0 35 5 3 2 Sample Carrier Recommended sample carriers for the Operetta Plate type Bottom mate...

Страница 60: ...ured 5 3 3 Power Supply The Operetta is connected to mains by a separate power cable which is specified as follows Specifications Europe US Cordage Rating 16A 250V AC 15A 125V ACl Plug 16A 250V AC IEC...

Страница 61: ......

Страница 62: ...PerkinElmer Schnackenburgallee 114 22525 Hamburg Germany Internet http www perkinelmer com email info perkinelmer com...

Отзывы: