Some objectives also feature a secondary scale, which does not contain the numerical
values. In the case of the “20x WD” lens this secondary scale consists of a single
mark at the 1 mm value. Below is the secondary scale for a 0.75 mm adjustment is
shown.
How to Adjust the Correction Collar
Objective with correction collar
1. If the message
"Please adjust the correction collar to ... mm"
appears, open the
sample lid.
2. Turn the correction collar and adjust it to the given value [mm].
3. If the mark of the correction collar is not visible from your position, use the
provided mirror to adjust for the correct value.
3 Retooling and Maintenance
Operetta Technical Manual
37