Each newly added point of the calibration curve is assigned with a number (1), 2), 3) and so on).
The max number of the calibration curve points
– 128.
Note.
If calibration is carried out on any of the listed calibration blocks in pos. 4, Figure
30, then an operator can easily activate marker STD (pos.3, Fig. 30) and select an appropriate
type of the calibration block before adding of the calibration curve points. Then add in order,
starting with a minimum, all signals from defects of a given calibration block and the flaw
detector will automatically assign the wanted depth to the defects.
4.14
Communication with PC
The flaw detector in a set with "PCE-USC 30"
software ensures communication mode with a PC
for the information input to the PC from the flaw detector memory (testing results, operating
setups) and the possibility to print out this information and to form a report.
4.14.1
Running the Program
To run the program it is required:
- to copy the files to the PC hard disk drive from the disk, that is included to the flaw detector
delivery set;
- run the
file “.exe”.
4.14.2
Program Description
The operating window of the program includes the following areas (Fig. 32):
1 - main menu; 2 - tab panel; 3 - time diagrams area; 4
– signal display area (active area); 5 –
bottom status bar
Fig. 32
– Active window of PCE-USC 30 program