After pressing
button, there appears a window to preview the testing protocol.
There is an option to select between: weather to print out the setting parameters of the flaw
detector or not.
To abstain from printing out the flaw detector settings, put a tick in front of
"Disable setting".
To print-out the testing protocol, press
"Print"
button in the upper part of a testing
protocol window.
To save the testing protocol in a «*.pdf» format, press
button.
To exit the review mode of a testing protocol it is required to press
button.
4.14.2.2
Tab Panel
Tab panel includes three tabs. Each tab corresponds to the selected frequency or to the mix of
frequencies (MIX): Frequency 1, Frequency 2, MIX.
Note
- Frequency 2 and MIX are available for viewing only if the testing is carried out in a
two-frequency mode
.
4.14.2.3
Time diagram area
Two diagrams of signal to time dependence are displayed in time diagram area (item 3, Fig. 32).
4.14.2.4
Signal display area (active area)
In the display area of eddy current signals is displayed the eddy current signal, which is
represented by an image in the complex plane for the frequency selected (item 4, Fig. 32). In
the active area are displayed the signals that come into the area covered by the measuring
cursor.
4.14.2.5
Bottom status bar
Fig. 37
"Duration of time sweeps"
– sets the number of measurements to be displayed on time
sweeps. Duration of time sweep is measured in a range from 256 to 32768.
“Type of measurement”
– allows to select the required type of cursor measurement from the
following list:
-
Peak-Peak
(the measurement is made between two points situated at the max. distance from
each other in cursor width);
-
Vrt-M
(vertical maximum
– the measurement is made between two points situated at the max.
distance from each other in cursor width in vertical projection);
-
Cnt-Pk
(centre peak - the measurements is made between the cursor centre and the point
situated at the max. distance from the centre in cursor width);
-
Cnt-K
(centre cursor - the measurements is made between the centre and the cross point of
cursor centre signal);