© PCE Instruments
42
Eng
li
s
h
4.11
“Results” Menu
The given menu allows storing and loading of the testing results.
In selecting "Results" menu the flaw detector displays the window with a list of results (Figure
26). The list of results contains catalogues (folders) and files. The catalogues are displayed in
the top part of the list (and are marked with the folders icon), the files
– in the bottom part and
are numbered. The catalogues, in their turn, can comprise subdirectories and files.
A new catalogue is created by placing the cursor on the line with colon or on the existing
catalogue and pressing
key ("F2 Create"). The existing catalogue is deleted by pressing
key ("F3 Delete").
Having placed the cursor on an empty string or on the existing file, function keys change their
function for the operation with files of results or settings.
Fig. 26
– "RESULTS" menu. List of catalogues and files
Note
. To prevent an accidental deleting of catalogues with useful information, the flaw
detector has a protection, which allows only empty catalogues deleting. In order to delete the
catalogues, firstly you need to delete its contents.
With the help of function keys
,
,
– "F1 Load", "F2 Save",
"F3 Delete" (see Figure 27)
– an operator can load the selected result (settings and data are
automatically loaded), save the current result and delete the selected result.
Fig. 27 - Operations with the testing results