36
0
TROUBLESHOOTING
WARNING
z
z
If the recliner is damaged in an earthquake, fire, flood or other incident,
contact us (see “Contact Information” on p. 46) or your distributor
for inspection and repair. Otherwise, there is a risk of electric shock
or fire due to short circuit or current leakage from electrical parts, or
injury due to malfunctioning caused by deformation of the recliner.
z
Problems may not always indicate faults. Check the following once again before requesting repair.
z
If the problem persists after taking corrective action, stop using the recliner immediately, disconnect
the power plug from the outlet and contact us (see “Contact Information” on p. 46) or your
distributor for repair.
z
In case of a failure or temporary problem, the display panel
on the handheld controller displays an error code.
Check the error code.
Status
(symptom)
POWER lamp
on handheld
controller
Error code
on handheld
controller
Check
Corrective action
Reference
page
The recliner
does not
move when
the operation
buttons are
pressed.
OFF
–
Is the power plug plugged into the
outlet?
Plug the power plug into the outlet.
–
–
Is the power plug plugged into the
outlet, but the POWER lamp on
the handheld controller is OFF?
Plug another electrical device into the outlet to
check that power is available.
–
Green lights
–
Is the operation selection button
for the pressed operation button
OFF?
Operate the operation disabling button to
enable operation buttons.
26
Slow flashing
in red only
when the
button is
pressed
(every
1 second).
U0
Is the power plug plugged into the
outlet while the operation button is
pressed?
Release the button that has been pressed.
–
U1
Is there any obstacle that is
interfering with the movable parts
of the recliner?
Remove the obstacle, then operate the
recliner.
–
U2
Are you using an electronic device
in the recliner, such as electronic
treatment apparatus?
It seems that a preventive action to detect a
poor communication condition was activated.
This is not an error. When using electronic
treatment apparatus, disconnect the power
plug for the recliner from the power outlet.
–
–
Unplug the power plug from the outlet and
check that the POWER lamp on the handheld
controller goes out, then plug the power plug
back into the outlet.
–
U3
–
Unplug the power plug from the outlet and
check that the POWER lamp on the handheld
controller goes out, then plug the power plug
back into the outlet.
–
U4
Did you use the recliner
continuously? (When the actuator
is operated continuously, the
operation time limit may be
activated to stop operation.)
Do not perform any operation on the recliner
for 20 to 57 minutes. (If you don’t observe
this no-operation time, the error code will be
displayed again).
–
U5
–
Unplug the power plug from the outlet and
check that the POWER lamp on the handheld
controller goes out, then plug the power plug
back into the outlet.
–
U9
Did you press the button for a
long period of time (30 seconds
or longer) while operation is
stopped?
Release the button and operate again.
–
–
Unplug the power plug from the outlet and
check that the POWER lamp on the handheld
controller goes out, then plug the power plug
back into the outlet.
–
Power lamp
Error code