44
Guidance and Manufacturer's Declaration — Electromagnetic Immunity
The recliner is intended for use in the electromagnetic environments specified below.
The customer or the user of the recliner must use it in an environment specified here.
Immunity test
IEC 60601-1-2 test level
Compliance level
Electromagnetic environment — guidance
RF conductivity
IEC 61000-4-6
3 V
0.15 MHz to 80 MHz
6 V
ISM bandwidth between
0.15 MHz and 80 MHz
80% amplitude modulation
(1 kHz)
3 V
0.15 MHz to 80 MHz
6 V
ISM bandwidth
between 0.15 MHz
and 80 MHz
80% amplitude
modulation (1 kHz)
Field strengths from fixed RF transmitters,
as determined by an electromagnetic site
survey, must be less than the compliance
level in each frequency range.
Interference may occur in the vicinity
of equipment marked with the following
symbol:
Radioactive RF
electromagnetic
field
IEC 61000-4-3
3 V/m
80 MHz to 2.7 GHz
80% amplitude modulation
(1 kHz)
3 V/m
80 MHz to 2.7 GHz
80% amplitude
modulation (1 kHz)
a. Field strengths from fixed transmitters, such as those of base stations for radio (cellular/cordless) telephones, land mobile
radios, amateur radios, AM and FM radio broadcasts, and TV broadcasts cannot be predicted theoretically with accuracy.
To determine the electromagnetic environment due to fixed RF transmitters, an electromagnetic survey of the site should be
considered. If the measured field strength in the location in which the recliner is used exceeds the applicable RF compliance
level above, the recliner should be observed to verify whether or not it is operating appropriately. If an abnormal operation is
seen, take additional measures as needed, such as changing the location of use of the recliner.
b. These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
structures, objects, and people.