CGA 351 User’s Manual
19
Chapter 3.
Configuration Options (cont.)
Note:
The
[0:Analog Output]
,
[2:Alarm-1]
,
[2:Alarm-2]
,
[2:Alarm-3]
and
[2:Alarm-4] options
are also available in the
above box.
To select the desired output(s) from the previous list box, complete the following steps:
1.
Use the [
] and [
] keys to highlight a specific output. For this example, make sure that
[0: Display]
is selected
as one of the outputs.
2.
Use the [
] and [
] keys to select
“Add” or “Remove” and press the
[ENTER]
key to change the status of the
selected output. An output is selected if an asterisk (*) appears to its left.
3.
When all of the desired outputs have been selected, use the [
] and [
] keys to select
“OK” and press the
[ENTER]
key.
Notice that the newly configured display shows the
slot number and the input parameter on the top line, and the
current
live value along with the unit type are displayed in the center of the screen.
Note:
See “The [Opt-User] Menu” on page 46 for instructions on how to adjust the display contrast.
Table 9:
From 1: Cfg Ox..pcnt
A list of the available output
devices is shown. Follow the
instructions below to select
the desired outputs.
*0: Display
*0: Fault Alarm
Add Remove
OK
Cancel
POWER
ESC
ENTER
Table 10:
1 Oxygen
Press
[ESC]
four times to
return the display to the
actual set configuration.
2.71
ppm
POWER
ESC
ENTER
Содержание CGA 351
Страница 1: ...panametrics com Flow 910 199 Rev D Aug 2021 CGA 351 Zirconium Oxide Oxygen Analyzer User s Manual...
Страница 2: ......
Страница 4: ...Warranty 4 CGA 351 User s Manual no content intended for this page...
Страница 32: ...CGA 351 User s Manual 26...
Страница 94: ...CGA 351 User s Manual 88...
Страница 98: ...Chapter 6 92 CGA 351 User s Manual no content intended for this page...
Страница 104: ...Appendix B 98 CGA 351 User s Manual...
Страница 112: ...CGA 351 User s Manual 106...
Страница 118: ......