Appendix C.
102
CGA 351 User’s Manual
C.1.4b Replacing the Fuses
If the main power fuses are defective, see Figure 9 below and install new fuses by completing the following steps:
1.
Disconnect the main power to the Rack Mount CGA 351.
2. Insert a screwdriver blade behind the fuse access door (to the right of the power switch) and pry it open.
3. Using needle-nose pliers, pull each of the two fuse holders (with an arrow printed on the end) straight out of the
unit.
4. Replace the fuses with new ones of the same type and rating. See Chapter 6,
Specifications, for the correct fuse
to use.
5. Reinsert the fuse holders, making sure the arrows are pointing down, and close the fuse access door.
The Rack Mount CGA 351 may now be placed back into service.
Figure 9: Rack Mount Fuse Locations
Fuse Access Door
Fuse Holder
Power Switch
Screwdriver
Fuse Holder
Power Connector