Chapter 4.
54
CGA 351 User’s Manual
4.1.9.6
Setting the Span Trim
Using the procedure under
Testing the Analog Output on page 52, set the Test Percent to 100.
At the analog output setup menu, proceed as follows:
Note:
The
[Aout Type], [Zero Setpoint]
and
[Span Setpoint]
options are also available in the above list box.
At the next prompt, enter the desired value for the span trim.
The meter now automatically returns to the “Analog Output” prompt.
Using the procedure under
Testing the Analog Output on page 52, reset the Test Percent to its normal value.
Either press
[ESC]
until you exit the
User Program or proceed to the appropriate section to continue setting up the
analog output.
Table 88:
User Aout
Use the [
] and [
] keys to
select
[Span Trim]
, and press
[ENTER]
.
Test
Zero Trim
Span Trim
POWER
ESC
ENTER
Table 89:
User Aout Span-Trim
Use [
] and [
] to increment
the value. Press
[ENTER]
when
done.
Span Trim
p
0 q
POWER
ESC
ENTER