Chapter 4.
48
CGA 351 User’s Manual
Setting Up the Analog Output (cont.)
The following procedures may be performed in this submenu:
•
Analog Output Type:
configure the analog output for a 0-20 mA, 4-20 mA, 0-2 V, or Namur type.
•
Zero Setpoint:
enter the live measurement value that corresponds to the low end of the analog output range.
•
Span Setpoint:
enter the live measurement value that corresponds to the high end of the analog output range
minus the zero setpoint.
•
Test:
specify a fixed percentage of the analog output range and verify that the correct value is generated.
•
Zero Trim:
adjust the actual output to equal the low end of the analog output range.
•
Span Trim:
adjust the actual output to equal the high end of the analog output range minus the zero setpoint.
Proceed to the appropriate sub-section to program the desired analog output feature.