Chapter 3.
20
CGA 351 User’s Manual
3.1.5.3
Using the Auto Range Feature
The Auto Range feature includes three
unit types:
•
ARng (Auto Range %):
the ratio of the current oxygen reading to the currently active Auto Range, expressed as a
percentage
•
AR-A (Auto Range Alarm A):
automatically set to a value of
0 or 1
•
AR-B (Auto Range Alarm B):
automatically set to a value of
0 or 1
As an example of how these units work with the default values listed in Table 64 on page 43, consider a situation
where the current oxygen reading is
7.5%
. The Auto Range feature would select the current active range as
Range 3
in
the table. Then,
This means that the current oxygen reading represents 75% of the currently active Auto Range. In addition to the
ARng value calculated above,
AR-A = 0
and
AR-B = 1
, as shown in the table.
After programming the Auto Ranges, the CGA 351 must be configured to properly handle the Auto Range output. To
accomplish this, the following outputs must be reprogrammed:
•
Alarm A and Alarm B
• 4-20 mA analog output
ARng
%O2
AR Zero
–
AR Span
AR Zero
–
---------------------------------------------------------- 100
75%
=
=
Содержание CGA 351
Страница 1: ...panametrics com Flow 910 199 Rev D Aug 2021 CGA 351 Zirconium Oxide Oxygen Analyzer User s Manual...
Страница 2: ......
Страница 4: ...Warranty 4 CGA 351 User s Manual no content intended for this page...
Страница 32: ...CGA 351 User s Manual 26...
Страница 94: ...CGA 351 User s Manual 88...
Страница 98: ...Chapter 6 92 CGA 351 User s Manual no content intended for this page...
Страница 104: ...Appendix B 98 CGA 351 User s Manual...
Страница 112: ...CGA 351 User s Manual 106...
Страница 118: ......