CGA 351 User’s Manual
99
Appendix C.
Appendix C.
C.1 Optional Enclosures
C.1.1 Introduction
Upon request, the
CGA 351 analyzer may be supplied in an enclosure other than the standard NEMA-4X enclosures
described in Chapter 1 of this manual,
General Information. Although the standard installation and wiring instructions
set forth in Chapter 2,
Installation, will still apply in general terms, some of the details may vary for different enclosure
types.
C.1.2 Rack Mount Installation
The CGA 351 analyzer is available in a rack mount enclosure for installation in a standard 19” electronics rack (see
Figure 6 below).
Figure 6: CGA 351 Rack Mount Version
C.1.2a Mounting the Rack Mount
Refer to Figure 10 on page 103 for mounting dimensions and install the rack mount CGA 351 as follows:
1.
Slide the unit into the rack at the desired height.
2. Fasten the unit securely to the rack with four screws (two on either side of the front panel).
After the rack mount CGA 351 has been physically mounted into the rack, proceed to the next section for connecting
the sample system.