CGA 351 User’s Manual
21
Chapter 3.
3.1.5.4
Configuring the Alarms
The first step in setting up the alarms is to assign the correct unit types. To accomplish this, proceed as follows from
the
Main Menu:
Note:
The
[ppm], [pcnt] and [AF]
options are also available above.
Table 11:
Cfg
Cal Opt Disp
Upon entering the
User
Program, the
[Cfg]
option is
already selected. Simply press
[ENTER]
.
POWER
ESC
ENTER
Table 12:
Cfg
Cal Opt Disp
To proceed with the
[1: Cfg]
option, press
[ENTER]
.
1: Cfg
POWER
ESC
ENTER
Table 13:
From 1: Cfg
A list of the available display
inputs is shown. Use the [
]
and [
] keys to select the
Oxygen input and press
[ENTER]
.
Oxygen
Temperature
Sensor
POWER
ESC
ENTER
Table 14:
From 1: Cfg Oxygen
A list of the available unit
types is shown. Use the [
]
and [
] keys to select the
AR-A option and press
[ENTER]
.
ARng
AR-A
AR-B
POWER
ESC
ENTER