CGA 351 User’s Manual
39
Chapter 4.
4.1.5.5
Selecting the Proportional Band
CAUTION!
Improper temperature controls settings can seriously degrade the performance of the instrument.
Never change these settings without specific instructions from the factory.
Enter the temperature control menu, as described on page 34, and proceed as follows:
Note:
See page 5 for a discussion of the proportional band.
Note:
The
[Setpoint], [Sensor Temp], [Int Time]
,
[Max.Duty Cycle]
and
[Default]
options are also available in the above
list box.
At the next prompt, enter the desired value for the proportional band.
The meter now automatically returns to the “Temp. Control” prompt. Either press
[ESC]
until you exit the
User Program
or proceed to the appropriate section to continue setting up the temperature control.
Table 53:
Temp-Cntrl
Use the [
] and [
] keys to
select
[Prop Band]
and press
[ENTER]
.
Air Offset
Select Gas Type
Prop Band
POWER
ESC
ENTER
Table 54:
Temp-Cntrl Se
. .
Prop
Using the [
] and [
] keys,
position the cursor under the
desired character. Then, use
the [
] and [
] keys to
increment the value. When
done, press
[ENTER]
.
Prop
0000
p
POWER
ESC
ENTER
Table 55:
Tem
. .
Unit-measuring
This message indicates that
changes in this menu will alter
the live data output. To
accept the new value, use the
[
] and [
] keys to select
[Yes]
and press
[ENTER]
.
Unit measuring
Accept changes?
Yes
No
Cancel
POWER
ESC
ENTER
Содержание CGA 351
Страница 1: ...panametrics com Flow 910 199 Rev D Aug 2021 CGA 351 Zirconium Oxide Oxygen Analyzer User s Manual...
Страница 2: ......
Страница 4: ...Warranty 4 CGA 351 User s Manual no content intended for this page...
Страница 32: ...CGA 351 User s Manual 26...
Страница 94: ...CGA 351 User s Manual 88...
Страница 98: ...Chapter 6 92 CGA 351 User s Manual no content intended for this page...
Страница 104: ...Appendix B 98 CGA 351 User s Manual...
Страница 112: ...CGA 351 User s Manual 106...
Страница 118: ......