CGA 351 User’s Manual
53
Chapter 4.
4.1.9.5
Setting the Zero Trim
Using the procedure under
Testing the Analog Output on page 52, set the Test Percent to 0.
At the analog output setup menu, proceed as follows:
Note:
The
[Aout Type], [Zero Setpoint]
and
[Span Trim]
options are also available in the above list box.
At the next prompt, enter the desired value for the zero trim.
The meter now automatically returns to the “Analog Output” prompt.
Using the procedure under
Testing the Analog Output on page 52, reset the Test Percent to its normal value.
Either press
[ESC]
until you exit the
User Program or proceed to the appropriate section to continue setting up the
analog output.
Table 86:
User Aout
Use the [
] and [
] keys to
select
[Zero Trim]
, and press
[ENTER]
.
Test
Zero Trim
Span Trim
POWER
ESC
ENTER
Table 87:
User Aout Zero-Trim
Use [
] and [
] to increment
the value. Press
[ENTER]
when
done.
Zero Trim
p
0 q
POWER
ESC
ENTER