Autodos M
MA60-10E
ENGLISH
Autodos M
MA60-10E
LEDs – indicators for calibration in progress
Active
Calibration in progress; calibration of pH7, pH9 and chlorine zero points possible.
Adjust
Calibration in progress; calibration of chlorine values possible.
If the “
Active
” LED flashes, it indicates you are in configuration mode (see the chapter System configuration on page 21).
To exit configuration mode, press and hold the
Cal.
button [0] until the “
Active
” LED stops flashing.
LEDs – indicators for other functions
Code
Indicates code mode (change of settings possible), steady = authorization level 1, flashing = authorization level 2.
Stdby
Indicates input that switches off alarm and dosing; the alarm relay is activated when the input is activated.
Text on the display
Runtime texts
A list of texts that may be displayed during normal program operation and the contexts in which they occur is found below:
- - - EEP Er1
at first start-up or EEPROM error; reset gives default setting.
dEL SEC # # #
restart performed, # # # indicates the time before start delay is finished.
Error codes
See section 12: Troubleshoting, sid 32.
Start-up sequence
When Autodos M is started up, a text (see below) is shown. The three digits in the text indicate how many seconds are left until
the end of the start-up sequence. The waiting time during start-up is intended to avoid problems with alarms being triggered
because the measurement values have not had time to stabilize, etc. (The start delay can be bypassed by pressing and holding
the
Reset
button [15] for about one second.)
Autodos M3
:
”dEL SEC 059”
Autodos M2
:
”SEC 059”
Autodos M1
: ”059”