Required accessories:
A) Isopropyl Alcohol (91% or better)
B) Round
wooden
toothpick with sharp point
(NOTE: do not use a metal pin or needle since metal will scratch the
surface of the detector)
C) Cotton swab
D) Jeweler’s loupe (10x magnification recommended)
E) Compressed Air (not shown)
Place a small amount of cotton from the swab onto the wooden
toothpick as shown at right.
Wet the cotton toothpick tip with the isopropyl alcohol.
Remove the adapter cap from the power meter detector port.
USING GREAT CARE
, gently insert the cotton toothpick tip into the detector port.
Using as little pressure as possible, rotate the cotton toothpick tip in the detector port
to clear away any dust or dirt that has accumulated on the detector.
Using light pressure from the compressed air, blow out the detector port.
Use the jewelers loupe to inspect the detector end face.
If dust or dirt are still present, repeat steps
until the detector port is free of debris.
If no dust or dirt is found, replace the adapter cap on the detector port and cover the detector
port with the dust cap.
WARNING! BE VERY CAREFUL WHEN INSERTING THE
TOOTHPICK INTO THE DETECTOR PORT. THERE IS A VERY THIN
GLASS WINDOW THAT WILL EASILY BREAK WITH TOO MUCH
PRESSURE.
OPERATION/MAINTENANCE
CLEANING THE DETECTOR PORT
A
B
C
D
56