OPERATION/MAINTENANCE
SETUP MENU
SETUP MENU > POWER OPTIONS
-- Set duration for various power saving modes
DIM TIME
minutes before display dims
Range of values: 1 to 250
STANDBY TIME
minutes before unit goes into standby mode (only display turns off)
Range of values: 2 to 250
OFF TIME
minutes before unit turns completely off
Range of values: 3 to 250
SAVE
Save settings and return to SETUP MENU
EXIT
Exit without saving settings and return to SETUP MENU
HELP
Context-sensitive help
SETUP MENU > CUSTOM STANDARD
-- configure a custom standard with user-definable standard parameters. Parameters are
set for each fiber type.
FIBER TYPE
Options: 62.5um OM1 / 50.0um OM2 / 50.0um OM3 / 50.0um OM4
Indoor SM / Outdoor SM
Select a fiber type and press EDIT to set parameters for that fiber type.
Wave (nm)
Up to 2 wavelengths (column A and B) can be defined per fiber type
Options: 850, 980, 1300, 1310, 1490, 1550, 1625, 0
Two types of standards can be defined:
GENERIC STANDARDS (link budget is calculated based on link configuration)
uses Loss/km, Max len, Con loss, and Spl loss
APPLICATION STANDARDS (link budget is a fixed number)
uses Max loss, Min loss, and Max len
GENERIC STANDARD PARAMETERS
Loss/km (dB)
Amount of acceptable dB loss (attenuation) per kilometer of fiber
Max len (m)
Maximum length of fiber link in meters
Con loss (dB)
dB loss per interconnection; defines both reference-grade and standard-grade reference cable connectors
Spl loss (dB)
dB loss per splice; can be either fusion or mechanical splices
APPLICATION STANDARD PARAMETERS
Max loss (dB)
loss measurements that exceed Max loss will show as a FAIL
Min loss (dB)
loss measurements that do not reach the Min loss will show as a FAIL
Max len (m)
Maximum length of fiber link in meters
52