SET REFERENCE
TEST PROCEDURES
35
After a few seconds, the SOURCE REFERENCE POWER will be shown. Press DONE when all
references have been set.
Set the light source to the wavelength shown or AUTO mode if the source is an OWL source that is
capable of AUTO mode (which allows simultaneous dual-wavelength testing).
Press TEST to set the SOURCE REFERENCE POWER.
Optical Source
WaveSource Pro
AUTO
PORT
MOD
CW
Wavelength
Set source to AUTO or 1310nm
You will need to set a source
power reference for each
wavelength that you are testing.
Press TEST to measure.
Press DONE after testing all
wavelengths.
Power
Typical
TEST
WAVE
BACK
DONE
SOURCE REFERENCE POWER
HELP
Wavelength
1310nm
1550nm
Set source to AUTO or 1310nm
You will need to set a source
power reference for each
wavelength that you are testing.
Press TEST to measure.
Press DONE after testing all
wavelengths.
TEST
WAVE
BACK
DONE
SOURCE REFERENCE POWER
HELP
Power
-10.21dBm
-10.60dBm
Typical
-10.00dBm
-10.00dBm
Optical Source
WaveSource Pro
AUTO
PORT
MOD
CW
NOTE: if not using an AUTO
source and a second wavelength
is required, connect a separate
reference cable between the
testers and follow the procedure
for the second wavelength.
LOSS MODE