TEST PROCEDURES
PATCH CABLE VERIFICATION
OPM MODE
Check patch cables before testing to determine if they are okay to use.
NOTE: when using a F7L (Fiber OWL 7 LITE) power meter, a separate light source is required for checking patch cables.
Connect a patch cable between the OPM port and the source port on the Fiber OWL 7.
NOTE: If AUTO mode was chosen, both wavelengths will fill in automatically.
If CW mode was chosen, press WAVE so that the OPM wavelength is the same as the chosen source wavelength.
The patch cable measurement should be close to the NIST calibrated power level:
(may be a little higher or lower)
SINGLEMODE (1310/1550): -10.00 dBm
62.5/125um MULTIMODE (850/1300): -20.00 dBm
50/125um MULTIMODE (850/1300): -23.00 dBm
NOTE: consider replacing patch cables that fall more than 1 dB below these suggested values