SET REFERENCE
TEST PROCEDURES
21
CERT MODE
After a few seconds, the REFERENCE CABLE LOSS will be shown. (NOTE: If Loss exceeds Maximum,
it is recommended to replace the additional test cable and/or mating sleeve.)
Press DONE to record the REFERENCE CABLE LOSS and continue.
Optical Source
WaveSource Pro
AUTO
PORT
MOD
CW
TEST
WAVE
BACK
DONE
REFERENCE CABLE LOSS
HELP
Wavelength
1310nm
1550nm
Set source to AUTO or 1310nm
You will need to set a source
power reference for each
wavelength that you are testing.
Press TEST to measure.
Press DONE after testing all
wavelengths.
Loss
0.21dB
0.40dB
Maximum
0.75dB
0.75dB
Press TEST to measure the REFERENCE CABLE LOSS.
Optical Source
WaveSource Pro
AUTO
PORT
MOD
CW
TEST
WAVE
BACK
DONE
REFERENCE CABLE LOSS
HELP
Wavelength
Set source to AUTO or 1310nm
You will need to set a source
power reference for each
wavelength that you are testing.
Press TEST to measure.
Press DONE after testing all
wavelengths.
Loss
Maximum