CONFIGURE JOB PARAMETERS
TEST PROCEDURES
34
On the METER, YES to begin the SET
REFERENCE procedure.
Power ON the light source to be used
for the test.
Setting a reference
(aka “zeroing”)
Sets a baseline optical power
level between a power meter
and light source prior to
taking measurements.
The difference between this
reference level and the
link power measurement
is called optical loss.
You must set a zero reference
before testing. Do you wish
to do it now?
YES
NO
Connect the appropriate reference cable as shown between the OPM port on the Fiber OWL 7 and
the light source port being used for this test, then press any key on the meter to proceed.
Optical Source
WaveSource Pro
AUTO
PORT
MOD
CW
Using your test reference
cables, please connect the meter
and source as shown and
press any key to proceed.
Optical Source
WaveSource Pro
AUTO
PORT
MOD
CW
LOSS MODE