E-026-3-E

3

(8)  When  you  select  "Other  scale"  on  the  window  shown  in  Fig.  3-69,  the  input  of  other  scale  will  be 

enabled as shown in Fig. 3-73. If you enter an analog value (here, 1V) equivalent to 4mA output in the 
4mA entry field, the corresponding value will be automatically entered in the 20mA entry field.

Fig. 3-73

 

Here, analog output adjustment with 1 to 5V scale is available.

 

By clicking "OK" button, Fig. 3-70 will be displayed in the same manner. To execute the analog output 
adjustment,  click  "OK"  button.  Additionally,  click  "OK"  button  to  start  the  adjustment.  Connect  a 
measuring instrument (reference voltmeter) across both ends of load resistance R

L

.

(9)   The window shown in Fig. 3-74 will be displayed. Select whether 4mA adjustment or 20mA adjust-

ment should be executed.

 

Here, 4mA adjustment is executed and then 20mA adjustment is executed in order.

 

Enter the reading of a measuring instrument and click "OK" button.

Fig. 3-74

 

When  the  reading  is  sent,  the  transmitter  automatically  adjusts  the  output  to  4mA.  Verify  that  the 
indicator of the connected measuring instrument is 1V.

 

If  another  adjustment  is  required,  enter  the  reading  of  a  measuring  instrument  on  this  window  and 
click "OK" button.

(10) To execute 20mA adjustment, click 20mA on the window shown in Fig. 3-74 and execute adjustment 

so that the connected measuring instrument indicates 5V as in the same way as 4mA adjustment.

(11) By clicking "Cancel" button, the adjustment of analog output value will be completed. The window 

"Adjustment is completed" shown in Fig. 3-72 will be displayed. Click "OK" button.

(12) To stop the adjustment during the process, click "Cancel" button and follow the following message.

It is not necessary to enter 
a value in the lower field. 
If you enter a value in the 
upper field, the value will be 
automatically entered in the 
lower field.

Click here and select 
which adjustment 
should be executed.

Содержание EL2310-0DE Series

Страница 1: ...owmeter for Liquid Psonic L4 SMART COMMUNICATION UNIT For the installation of application software LinkTop and the interface driver refer to Ins No E 020IMC Smart Communication Unit MODEL EL2310 Softw...

Страница 2: ...s 21 3 6 1 Diameter inside internal diameter of measuring tube 22 3 6 2 Fld dev vars transmitter variable 22 3 6 2 1 Flow parameter setting related to flow rate 22 3 6 2 2 Fluid parameter setting rela...

Страница 3: ...wmeter 57 3 11 3 Input Errors 58 3 11 4 Error Status Display List 59 4 PRODUCT CODE EXPLANATION 60 5 GENERAL SPECIFICATIONS 61 This manual uses the precaution words NOTE CAUTION and WARNING as explain...

Страница 4: ...rocedure Manual Ins No E 026 1 E 1 1 SMART COMMUNICATION UNIT 1 1 General Described in this manual are the operating instructions to use the application software LinkTop of the Smart Communication Uni...

Страница 5: ...and Devices Equipment set up with associated equipment and devices is shown in Fig 2 1 Prepare a PC by yourself Application software LinkTop Procedure of connection 1 Connect the cables as shown in F...

Страница 6: ...on Interface A signal transmitter is provided for the communication signal conversion between the external serial port COM port of the PC and the flowmeter When using Bell202 communication the Bell202...

Страница 7: ...ommunications when waiting for receiving data IDLE When communications are interrupted CLOSE When communications are disconnected 3 The overview of each menu is shown in the table below Display item D...

Страница 8: ...om left of the PC screen and then click LinkTop for Psonic L4 E Ver1 0 0 in Program 2 The following window Fig 3 2 will be displayed and it is possible to verify the version of the software Fig 3 2 3...

Страница 9: ...s possible to change the setting of serial port of PC 1 Click File F from the menu and then click on Port Setting I Ctrl I Fig 3 4 2 The following window Fig 3 5 will be displayed Here select the seri...

Страница 10: ...7 4 After the selection is completed click OK button After clicking the button the value will be changed to the setting value you select and the window will disappear If you want to cancel the setting...

Страница 11: ...then click on Connect C F3 Click here 2 The message box Fig 3 9 will be displayed By clicking OK button the connection will be started By clicking Cancel button the connection operation will be cance...

Страница 12: ...f window Fig 3 12 Fig 3 11 The display of communication status display bar switches between SEND and RECV WAIT By clicking Cancel button the connection will be cancelled 4 If the connection is complet...

Страница 13: ...connect D Shift F3 2 The message box Fig 3 14 will be displayed By clicking OK button the connection between flowmeter and LinkTop will be terminated By clicking Cancel button the connection terminati...

Страница 14: ...confirm and the application software window will disappear from the desktop To abort the terminating process click on Cancel button Fig 3 16 Fig 3 17 3 4 Terminating LinkTop 1 To exit LinkTop click Fi...

Страница 15: ...menu can be used to confirm process value of flowmeter instantaneous flow rate flow velocity cumulative total in positive direction cumulative total in opposite direction analog output Gain1 to 4 cha...

Страница 16: ...e total positive direction Backward totalizer cumulative total opposite direction Analog output analog flow rate Gain 1 gain value pass 1 Gain 2 gain value pass 2 Gain 3 gain value pass 3 Gain 4 gain...

Страница 17: ...Active then a tool bar as shown in Fig 3 20 will be displayed underneath the menu Fig 3 20 Fig 3 21 2 Click the button on the tool bar or Chart and Recording Interval from the menu then the window sho...

Страница 18: ...om the menu then the window shown in Fig 3 22 will be displayed enabling chart display of percent flow When changing the scale of the vertical axis enter the preferred values to Max and Min in Scale t...

Страница 19: ...in value passes 1 to 4 and Reynolds number are operated in a same way 7 If you click the button on the tool bar or Chart and Recording Recording Start from the menu the following dialog box Fig 3 24 w...

Страница 20: ...rameter of transmitter can be changed or not is displayed 1 Click Process Variables V in the upper part of the screen and then click on Write protect Then the following window Fig 3 27 will be display...

Страница 21: ...If windows of Diag Service menu are displayed on the screen then it will not be possible to set the various parameters transmitter information and so on Close these windows before making settings For...

Страница 22: ...e 1 Click Setup S from the menu and then click on Flow from the drop down list of Fld dev vars 2 The following window Fig 3 30 will be displayed Fig 3 30 3 6 1 Diameter inside internal diameter of mea...

Страница 23: ...Flow direction Inflow direction measurement mode Forward Flow Reverse Flow Both Total unit Totalization unit liter hliter CuMtr CuIn CuFt gal ImpGal bbl g kg MetTon ShTon LTon lb special Total decimal...

Страница 24: ...ring the steps 2 to 4 3 6 2 2 Fluid parameter setting related to fluid 1 Click Setup S from the menu and then click on Fluid from the drop down list of Fld dev vars 2 The following window Fig 3 33 wil...

Страница 25: ...k OK button Then the message box Fig 3 34 will be displayed Click OK here to change the settings to the inputted values and to reflect the changed setting values in the flowmeter s output If the flowm...

Страница 26: ...p to manual control so that the control loop cannot be influenced by the flowmeter s output 3 6 3 Outputs output setting 3 6 3 1 Pulse output pulse output setting 1 Click Setup S from the menu and the...

Страница 27: ...mode None no function contact output OFF Alarm ACTIVE when alarm is generated Zero adjust ACTIVE during zero point adjustment Flow detection ACTIVE at the occurrence of reverse flow None Alarm Zero a...

Страница 28: ...fety protocol change the control loop to manual control so that the control loop cannot be influenced by the flowmeter s output Fig 3 42 5 Click OK button If the value is changed to the selected setti...

Страница 29: ...menu and then click on Device information 2 The following window Fig 3 44 will be displayed Here set the transmitter information Fig 3 44 3 Set each item If you point the cursor to the input field the...

Страница 30: ...ch flowmeter 0 to 63 Loop current mode Loop current mode Disabled Enabled Num req preambles HART communication transmission preamble number Only for reference Num resp preambles HART communication rec...

Страница 31: ...t cumulative totals Close all Setup S windows to enable checking and adjustment functions Also note that other than Loop test the Diag Service T window cannot be displayed simultaneously with multiple...

Страница 32: ...ced by the flowmeter s output NOTE If the loop current mode is set to Disabled this test is unusable 1 Click Diag Service T from the menu and then click on Fix Analog from the drop down list of Loop T...

Страница 33: ...nalog value you set will be output While the simulated output value is output a message Simulated output is provided now is displayed at the bottom of the window as shown in Fig 3 51 Click Stop button...

Страница 34: ...be displayed Select the simulated output value and click Start button To set and output a desired output value select Other and enter the pulse frequency to be output then click Start button The avail...

Страница 35: ...53 Click Cancel button to exit Fix Pulse 3 7 1 3 Fix Flow simulated flow rate loop test This puts status output into a simulated output state and conducts a loop test on the output line Analog output...

Страница 36: ...he simulated output is created Fig 3 57 Setting item Detail Setting range Fix Flow Fixed output of instantaneous flow rate 9999 0 to 9999 0 setting unit 3 By clicking Start button the message box Fig...

Страница 37: ...ction of flowmeter transmitter 3 7 2 1 Auto zero auto zero point adjustment This adjusts the zero point for flow rate measurement of flowmeter transmitter 1 Click Diag Service T from the menu and then...

Страница 38: ...yed By clicking OK button the window will return to the window shown in Fig 3 61 Click Cancel button to exit it During this time the transmitter executes the zero point adjustment Fig 3 63 CAUTION The...

Страница 39: ...tment displays the process during adjustment as shown in Fig 3 66 3 7 2 2 Init AGC sensitivity adjustment This adjusts the sensitivity of flow rate sensor 1 Click Diag Service T from the menu and then...

Страница 40: ...K button The value after the adjustment will be displayed as shown in Fig 3 68 Fig 3 67 Fig 3 68 6 By clicking OK button the window will return to the window shown in Fig 3 64 Click Cancel button to e...

Страница 41: ...Click Diag Service T from the menu and then click on Trim Analog 2 The following window Fig 3 69 will be displayed Select the scale from 4mA 20mA or Other scale and click OK 3 By clicking OK button t...

Страница 42: ...ammeter indicates 4mA If another adjustment is required enter the reading of the ammeter again on this window and click OK button 5 To execute 20mA adjustment click 20mA side on the window shown in F...

Страница 43: ...ted in order Enter the reading of a measuring instrument and click OK button Fig 3 74 When the reading is sent the transmitter automatically adjusts the output to 4mA Verify that the indicator of the...

Страница 44: ...d 3 On this window Start Stop and Reset of cumulative total can be executed As Start Stop and Reset are interlocked with pulse output note that pulse is not output in Stop status In usual status Start...

Страница 45: ...26 3 E 45 3 8 Menu Maintenance maintenance Maintenance M is used for resetting rebooting the transmitter The actual screen is as shown in Fig 3 76 Fig 3 76 Maintenance Device reset Maintenance Item Tr...

Страница 46: ...clicking OK button the reset of transmitter will be completed 3 8 1 Device reset transmitter reset 1 Click Maintenance M from the menu and then click on Device reset 2 The following window Fig 3 77 wi...

Страница 47: ...rrently displayed windows For the types to be arranged Cascade TileHorizontal and TileVertical are available For instance assume that multiple windows are being displayed as shown in Fig 3 80 Fig 3 80...

Страница 48: ...3 E 48 3 If TileVertical is selected the windows are displayed horizontally as shown in Fig 3 83 Fig 3 82 Fig 3 83 2 If TileHorizontal is selected the windows are displayed vertically as shown in Fig...

Страница 49: ...or other types of storage medium Setting values can also be downloaded to the flowmeter transmitter However the data can be downloaded only when the write protect is set to Not write protect 3 10 1 1...

Страница 50: ...lected file can be printed For printing refer to 3 10 2 Printing 3 10 1 2 Saving Files It is possible to save parameters in the flowmeter transmitter to a file 1 Click File F from the menu and then cl...

Страница 51: ...operation anew Fig 3 86 3 10 1 3 Deleting Files It is possible to delete data files that are no longer necessary 1 Click File F from the menu and then click on Delete File X from the drop down list of...

Страница 52: ...suitable for this operation If you want to save the parameters follow 3 10 1 2 Saving Files Fig 3 87 4 After the click the selected file will be opened The message box Fig 3 88 will be displayed Click...

Страница 53: ...is stored from the drop down list 5 Click and select the file to be downloaded Verify File name N and click Open O button Click Cancel button to exit the operation without downloading the file Fig 3...

Страница 54: ...click OK button and the download is started the message box Fig 3 91 will be displayed It is possible to confirm the status of downloading by the progress bar in the center of the window 9 Once the d...

Страница 55: ...Fig 3 93 will be displayed To reflect the latest data execute uploading here If it is cancelled the printing screen will not be displayed Fig 3 93 3 After the upload is terminated the following windo...

Страница 56: ...y clicking Print button the printing will be executed as shown in Fig 3 95 5 For the printer to be used for printing carry out the setting by clicking Start menu Settings Control panel Printer of the...

Страница 57: ...kTop and Connections Is the smart communication unit s probe or unit disconnected Is the flowmeter transmitter receiving power Is the polling address correct Fig 3 96 3 11 2 If It is Not Possible to C...

Страница 58: ...u Diag Service Checking and Adjustment and the communications with flowmeter transmitter are tried an error box will appear on the desktop as an input error There are two types of input errors wrong s...

Страница 59: ...measurement Zero Offset Non Adjust Error Zero point not adjusted Implementation of zero point calibration Rom Check Error ROM check error Unable to recover automatically due to the hardware failure As...

Страница 60: ...Always 0 Applied flowmeter D Ultrasonic flowmeter for liquid Psonic L4 Language J Japanese Japanese version OS supported E English English version OS supported Interface 0 Without interface only appli...

Страница 61: ...ing of measurement value Reading change of setting and saving of parameter location of saving the file is FD HD or other external memory etc Adjustment of analog output Verification of loop of analog...

Страница 62: ...E 026 3 E All specifications are subject to change without notice for improvement 2017 04 Revised 2016 08 Released E 026 3 E 1...

Отзывы: