E-026-3-E
3
Fig. 3-62
(3) Click "OK" button. By clicking it, the message box (Fig. 3-62) will be displayed.
(4) By clicking "OK" button, the zero point adjustment will be executed. When it is started, the message
box (Fig. 3-63) will be displayed. By clicking "OK" button, the window will return to the window shown
in Fig. 3-61. Click "Cancel" button to exit it. During this time, the transmitter executes the zero point
adjustment.
Fig. 3-63
CAUTION: The process fluid in the measurement pipe of flowmeter has to be
completely stopped.
Otherwise, the zero point adjustment cannot be correctly executed.
Also, It may not be executed if an error exists in the flow rate measure-
ment.