E-026-3-E
3
Fig. 3-55
(4) By clicking "OK" button, the simulated output value you set will be output. While the simulated output
value is output, a message "Simulated output is provided now." is displayed at the bottom of the
window as shown in Fig. 3-55.
Click "Stop" button to stop the simulated output.
Fig. 3-56
(5) By clicking "Stop" button, the message box (Fig. 3-56) will be displayed. Click "OK" button.
(6) By clicking "OK" button, the window will return to the window shown in Fig. 3-53. Click "Cancel"
button to exit "Fix Pulse".
3.7.1.3 Fix Flow (simulated flow rate loop test)
This puts status output into a simulated output state, and conducts a loop test on the output line.
Analog output and pulse output will be created according to the simulated flow rate you set.
This test creates simulated output regardless of the process state. If the flowmeter's output is used to
control valves or other such parts, change the control loop to manual control for the sake of safety, and
so that the control loop cannot be influenced by the flowmeter's output.
(1) Click "Diag/Service (T)" from the menu, and then click on "Fix Flow" from the drop-down list of "Loop
Test".