(3) Click "OK" button. By clicking it, the message box (Fig. 3-65) will be displayed.
(4) By clicking "OK" button, the sensitivity adjustment will be executed. The screen during adjustment
displays the process during adjustment as shown in Fig. 3-66.
3.7.2.2 Init AGC (sensitivity adjustment)
This adjusts the sensitivity of flow rate sensor.
(1) Click "Diag/Service (T)" from the menu, and then click on "Init AGC" from the down-list of
"Calibration".
(2) The following window (Fig. 3-64) will be displayed. Stop the fluid completely, and then click "OK"
button.