(5) By clicking "Stop" button, the message box (Fig. 3-60) will be displayed. Click "OK" button.
(6) By clicking "OK" button, the window will return to the window shown in Fig. 3-57. Click "Cancel"
button to exit "Fix Flow".
3.7.2 Calibration (transmitter's adjustment function)
This executes the adjustment function of flowmeter transmitter.
3.7.2.1 Auto zero (auto zero point adjustment)
This adjusts the zero point for flow rate measurement of flowmeter transmitter.
(1) Click "Diag/Service (T)" from the menu, and then click on "Auto zero" from the drop-down list of
"Calibration".
(2) The following window (Fig. 3-61) will be displayed. Set the calibration time and stop the fluid
completely, and then click "OK" button.
Fig. 3-61
Setting item
Detail
Setting range
Calibration Dt time Zero point adjustment execution time
60 to 600 (seconds)