Extended Range
This function is primarily for safety purposes. It ensures that the maximum
SPL presented to the patient will not exceed approximately 110 dB SPL,
unless separately enabled by the operator.
# 9
This button enables presentation of high level intensities. When enabled, the
Ext. Range LED on the display panel will light up.
Change of frequency or function automatically disables the extended range.
Preprogrammed tests cannot contain intensity settings beyond the limits
defined by the Ext. Range function.
Without the Ext. Range function enabled, the maximum intensities in dB HL
for the various transducers are listed below:
Hz
Phone /
Bone
F.F.
Insert Phone
125
70
-
80
250
90
45
80
500
100
60
80
750
100
60
80
1000
100
70
80
1500
100
70
80
2000
100
70
80
3000
100
70
80
4000
100
60
80
6000
100
55
80
8000
80
-
80
When the audiometer is busy printing out to an external printer or to ZODIAC
901, the Ext. Range LED on the display panel will flash constantly (not
applicable in version 3.2 and higher).
Ext. Range LED
Flashes During
Printout
5.1.2 PATIENT COMMUNICATION & MONITORING CONTROLS
Talk Over
Pressing this button accesses the Talk-Over Mode interrupting the present-
ation of stimulus and permitting communication with the patient.
# 16
This pushbutton also displays the Talk-Over Level in dB for the purposes of
adjustment.
The display will show the level presented to the patient directly in dB HL and
this level may be changed by turning any of the rotary knobs.
Adjust the Talk-Over Level so that it is comfortably audible for the patient.
The selected level will be stored, even after the audiometer has been
powered off.
MIDIMATE 622 OPERATION MANUAL
5. Operating in Test Mode
23
Содержание MADSEN Midimate 622
Страница 1: ...Midimate 622 Clinical Diagnostic Audiometer User Manual Doc no 7 25 9000 09 Part No 7 25 900 0459 ...
Страница 16: ...10 MIDIMATE 622 OPERATION MANUAL 2 Installing Midimate 622 ...
Страница 18: ...Fig 3 The Selection Tree MIDIMATE 622 OPERATION MANUAL 3 General Description 12 Menu Menu Menu ...
Страница 54: ...This page has intentionally been left blank MIDIMATE 622 OPERATION MANUAL 7 Speech Audiometry 48 ...
Страница 70: ...This page has been intentionally left blank MIDIMATE 622 OPERATION MANUAL 10 Power On Options 64 ...
Страница 74: ...This page has intentionally been left blank MIDIMATE 622 OPERATION MANUAL 12 Calibration 68 ...
Страница 101: ...ZODIAC 901 Connection To 74 ZODIAC 901 Printout To 59 73 MIDIMATE 622 OPERATION MANUAL Subject Index 95 ...